IIC -407   MICROPROCESSOR BASED DIGITAL Ph METER
Technical Parameters
Function pH& temp. OR mV & temp, simultaneously
Display 16 character x 2 lines Alphanumeric BL LCD
Power supply 230 V ± 10% 50 Hz AC
PH
Range 0to 14 pH
Resolution 0.01 pH
Accuracy ± 0.01 pH ± 1 digit
Slope adjustment 80% to 120% By default 100%
Temp.compensation 0 to 100 ?C Auto with temp probe / Manual with keys
Auto buffer recognition 4.00 pH, 7.00 pH,9.20 pH
Calibration Auto& Manual, 3 point& 2 point (Programmable)
mV
Range -1999 mV to+1999 mV
Resolution 1 mV
Accuracy ± 0.1% ±1 digit
Calibration Two point manual
Temperature Range -20.07C to 2007C
Resolution 0.1 ?C
Accuracy ± 0.1% of range ±1 digit
Calibration two point manual
Sensors Pt100 probe
Data entry By six soft touch keys with audible sound
Storaqe 1000 samples
Real time display 24 Hour mode with Date
Interface RS-232, Centronics & computer
Print format All Data & Date Wise (Programmable)
Digital output Serial RS-232 Parallel Centronics (Programmable)
Error Indication pH ERROR.ORP ERROR.CAL ERROR & TERROR
Accessories pH electrode Instruction Manual Buffer Solutions
Dust cover Electrode Stand Temp Probe

Appearance & Specification subjected to change due to continuous development Print Format:-
DATE TIME TEMP(DegC) PARAMETER MODE SRNo
12/12/08 11:18 T= 020.0 pH= 07.00 MANUAL 001
12/12/08 11:18 T= 020.0 mV= 0000 MANUAL 002
12/12/08 11:18 T= 020.0 T= O2O.O0C MANUAL 003
DATE TIME TEMP(DegC) PARAMETER MODE SRNo
12/12/08 11:18 T= 020.0 pH= 07.00 MANUAL 001
12/12/08 11:18 T= 020.0 mV= 0000 MANUAL 002
12/12/08 11:18 T= 020.0 T= O2O.O0C MANUAL 003



IIC - 408   MICROPROCESSOR UV-VIS SPECTROPHOTOMETER DOUBLE BEAM (Variable Bandwidth) (Eight Cell Holder)
DOUBLE BEAM UV-VIS Spectrophotometer with more accuracy and flexible requirements. The two detectors are used to measure sample and refrence respectively and simultaneously for optimizing meaaurement accuracy It has vide wavelength range satisfying requirement of various fields. Such as biochemical research and industry. Pharmaceuticals analysis and production, education, environment protection. Food industry etc

SAILENT FEATURES
  • Wide Wavelength range, satisfying requirements of various fields.
  • Fully automated design, realizing the simplest measurement &satisfying the requirement of pharmacoppeia.
  • Maximum of 9 Wavelength & 8 sample can be measured at one time
  • Automatic change-over between W lamp &D2 lamp
  • Optimized optics and large scale integrated circuits, light source and receiver from world famous measurement methods all add up to high performance and reliability.
  • Rich measurement methods wavelength scan, time scan, multi-wavelength determination multi-order derivative determination, double w-wavelength method and triple-wavelength methods etc, meet difference measurement.
  • Automatic 10 mm 8-cell holder
  • Data Output can be optained via a printer port and a USB interface.
  • Parametters and data can be saved for user 5 convenience.
  • PC controller measurement can be achieved for more accurate and flexible requirement
Tecnnical Parameters
Optical System Double beam(1200 lines/mm Grating)
Wavelenqth Range 190 1100nm
Mode Basic / Quantitative / Wavelenqth Scan / DNA Protien Test/Kinetics
Scanning Speed Fast / Medium / Low
Band width 0.5/1.0/2.0/40 nm
Wavelength Accuracy ±0.3nm
Wavelength Repeatability 0.2nm
Photometric Accuracy ± 0.3 % T
Photometric Repeatability 0.2 % T
Photometric Display Range 0-200% T, 0.3-3 0 A, 0-9999C
Stability 0.001 A/h @ 500nm
Baseline Flatness ± 0.001 A
Noise ± 0.001 A
Stray Light <0.05 %T @ 220 nm, 360nm
Date Output Port USB
Printer port Parallel Port
Display Graphic LCD (320 240 Dots)
Lamps Deutenem Lamp & Tungsten Halogeon Lamp
Detector Silicon Photodiodes
Packing Dimension 860 x 660 x 465 mm (L x W x H)
Net Weight 26 Kg
Basic Mode To measure the Absorbance and transmiitance

Quantitative
  • Coefficient Method : Standard Curve Up to 10 Standard sample may be used to establish a curve. Four method for fitting a curve through the calibration points: Linear fit, Linear fit through zero. Square fit and cubic fit.
  • DNA/Protein Test
  • Concentration and DNA purity are quickly and easily calculated. Absorbance ratios 260 nm/280 nm with optional subtracted absorbance at 320 nm. DVA concentration^ 62.9xA260-36.0 x A280 protein concentration^ 552xA260-757.3xA280
Wavelength Scan
  • The wavelength scan intervals are 0.1,0.2,0.5,1,2,5 nm
  • Hi, Medium and low scan speed are available. They vary from 100 to 3600nm/min
  • Wavelength are scanned from high to low so that the instrument waits at high WL. and it minimizes the degradation of UV sensitive sample.
Kinetics
This mode may be used for time course scanning or reaction are calculations. Abs vs time graphs is displayed on the screen in real time Wait time and measurement time up to 12 hours may be entered with time interval of 0.5,1, 2,5, 10,30 seconds and one min. Post-run manipulation includes re-scalling, curve tracking and selection of the part of the curve required for rate calculation. Rate is calculated using a linear regression algorithm before multipling be the entered factor.